Hall Probes FH 54 + FH 55

Data Sheet PDF-DokumentDatenblatt PDF-Dokument

Transverse Probes for FH 54 + FH 55

Photo: Transverse Probes for FH 54 + FH 55

Transverse Probes for FH 54 + FH 55

Model HS-TGB5-104005 HS-TGB5-104010 HS-TGB5-104020 HS-TND5-0513026
W [mm] 4,0 max. 4,0 max. 4,0 max. 1,3 nom.
T [mm] (max.) 1,0 1,0 1,0 0,51
L [mm] (nom.) 55 100 200 26
A [mm] 2 ± 0,1 2 ± 0,1 2 ± 0,1 1 ± 0,1
H [mm] (nom.) 70 70 70 37
Cable C [m] 1,5 1,5 1,5 1
Stem Material Fiberglas epoxy Kapton
Aktive Area,
nominal Diameter [mm]
0,4 0,4 0,4 0,64
Ranges, full Scale 3 mT to 3 T
Corrected Accuracy
(% of Reading, DC)
0,25 % to 3 T 0,25 % to 3 T 0,25 % to 3 T 0,5 % to 1,5 T
Temperature
coefficient
(maximum)
Sensitivity
% / °C
± 0,02 (T) ± 0,02 (T) ± 0,02 (T) -0,1
(T): Probe with sensor for temperature correction.
Operating Temperature range 0 °C to 75 °C.
Due to continuous process improvement, specifications are subject to change without notice.

Transverse Probes for FH 54 + FH 55

Photo: Axial Probes for FH 54 + FH 55

Axial Probes for FH 54 + FH 55

Model HS-AGB5-4805 HS-AGB5-4820
D [mm] (nom.) 4,8 4,8
L [mm] (nom.) 55 200
A [mm] (nom.) 0,27 0,27
H [mm] (nom.) 70 70
Cable C [m] 1,5 1,5
Stem Material Fiberglas epoxy
Aktive Area,
nominal Diameter [mm]
0,4 0,4
Ranges, full Scale 3 mT to 3 T
Corrected Accuracy
(% of Reading, DC)
0,25 % to 3 T 0,25 % to 3 T
Temperature
coefficient
(maximum)
Sensitivity
% / °C
± 0,02 (T) ± 0,02 (T)
(T): Probe with sensor for temperature correction.
Operating Temperature range 0 °C to 75 °C.
Due to continuous process improvement, specifications subject to change without notice.